close Exit

Workstation for Pipettes Calibration

SDKP DUAL Workstation for Pipettes Calibration
This image is for reference only.
SDKP DUAL Workstation for Pipettes CalibrationSDKP DUAL Workstation for Pipettes CalibrationSDKP DUAL Workstation for Pipettes Calibration
close Exit
This image is for reference only.
SDKP DUAL Workstation for Pipettes CalibrationSDKP DUAL Workstation for Pipettes CalibrationSDKP DUAL Workstation for Pipettes Calibration
Availability:
In stock
Estimated shipping date :
15 working days

Need it urgently? Call us to check real-time stock availability: +44 (0)1625 569569

Price:
11200.80
 GBP
with 20% VAT, shipping cost not included
Choose a variant:
SDKP DUAL Workstation for Pipettes Calibrationkeyboard_arrow_down
  • SDKP Workstation for Pipettes Calibration
  • SDKP DUAL Workstation for Pipettes Calibration

Pcs.

Add to basket
We are near you.
We have been manufacturing weighing devices since 1984.

RADWAG UK Limited
16 Macon Court, Herald Drive, Crewe, CW1 6EA
+44 01625 569569


Delivery only on the territory of United Kingdom
Optional paid installation and training service available - please contact us for details and pricing.
Code:
WL-101-0173

Datasheet

Delivery components
table for calibration,
anti-vibration table ×2,
probe for temperature,
humidity and pressure measurement,
probe for water temperature measurement,
sink for used water,
armrests,
LCD monitor,
PC with pre-installed software PIPETTES RADWAG,
power supply,
keyboard and mouse,
container for used water
Chamber packaging size
820×1410×1000 mm
Gross weight
250 kg
Length
1350 mm
Width
750 mm
Height
1100 mm
Number of workstations
2

Overview


SDKP DUAL Workstation for Pipettes Calibration

 
 
In order to ensure measurement traceability, meet all the requirements resulting from supervision over measuring equipment and to facilitate  the process of pipettes calibration RADWAG recommends effective and ergonomic solution for calibrating automatic pipettes conducted by a user - complete workstation for pipettes calibration.

Application:
 Calibration of automatic pipettes:

  • with fixed volume
  • with adjustable volume.

Unit contents:

  • table for calibration
  • anti-vibration table x2
  • probe for temperature, humidity and pressure measurement
  • probe for water temperature measurement
  • sink for used water
  • armrests
  • LCD monitor
  • PC with pre-installed software PIPETTES RADWAG
  • absorbtion pump
  • power supply
  • keyboard and mouse
  • container for used water


Adapters for pipettes calibration
Pipette calibration adapters are intended for calibration and verification of piston pipettes using gravimetric method. Use of the pipette calibration adapter reduces maximum capacity of the balance by the mass of the weighing vessel. The table presents the minimum reguirements in accordance with ISO 8655 standard. It also enables to select suitable adapter for pipette calibration and respective balance version.
 


Selected volume1
of apparatus under test

V

 

Resolution
 mg
Repeatability
and linearity mg
Recommended balance
1 µl ≤ V ≤ 10 µl 0,001 0,002
MYA 21.5Y PLUS
MYA 21.5Y
XA 6/21.5Y.M.A PLUS
XA 6/21.5Y.M.A
XA 6/21.5Y.M
 
10 µl ≤ V ≤ 100 µl 0,01 0,02
XA 82/220.5Y.A PLUS
XA 82/220.5Y.A
XA 82/220.5Y
XA 120/250.5Y.A PLUS
XA 120/250.5Y.A
XA 120/250.5Y
100 µl ≤ V ≤ 1000 µl 0,1 0,2
1 ml < V ≤ 10 ml 0,1 0,2
Adapter for pipette calibration  MY 11 XA 11 XA 17 XA 100 XA 17 XA 100
Maximum vessel capacity  11 ml 11 ml 17 ml 100 ml 17 ml 100 ml
Weighing vessel mass  ~ 10 g ~ 10 g ~ 11 g ~ 65 g ~ 11 g ~ 65 g

1 For practical purposes, the nominal volume may be used to choose the balance [Reference: ISO8655-6:2002, pg. 2].
If the standard uncertainty of measurement of the balance is known (e.g. from the balance calibration certificate), it may be used instead of the repeatability and linearity. The standard uncertainty of measurement shall not be more than two to three times the resolution [Reference: ISO8655-6:2002, pg. 2].