Availability:
In stock
Estimated shipping date :
15 working days

Need it urgently? Call us to check real-time stock availability: +44 (0)1625 569569

Price:
4572.00
 GBP
with 20% VAT, shipping cost not included
Choose a variant:
MA 50.5Y Moisture Analyzerkeyboard_arrow_down
  • MA 50/1.5Y Moisture Analyzer
    Max: 50g | d: 0,1mg
  • MA 50.5Y Moisture Analyzer
    Max: 50g | d: 1mg
  • MA 110.5Y Moisture Analyzer
    Max: 110g | d: 1mg
  • MA 210.5Y Moisture Analyzer
    Max: 210g | d: 1mg
  • MA 200/1.5Y Moisture Analyzer
    Max: 200g | d: 0,1mg
  • MA 50/1.5Y.WH Moisture Analyzer
    Max: 50g | d: 0,1mg
  • MA 50.5Y.WH Moisture Analyzer
    Max: 50g | d: 1mg
  • MA 110.5Y.WH Moisture Analyzer
    Max: 110g | d: 1mg
  • MA 210.5Y.WH Moisture Analyzer
    Max: 210g | d: 1mg
  • MA 200/1.5Y.WH Moisture Analyzer
    Max: 200g | d: 0,1mg
  • MA 50/1.5Y.NS Moisture Analyzer
    Max: 50g | d: 0,1mg
  • MA 50.5Y.NS Moisture Analyzer
    Max: 50g | d: 1mg
  • MA 110.5Y.NS Moisture Analyzer
    Max: 110g | d: 1mg
  • MA 210.5Y.NS Moisture Analyzer
    Max: 210g | d: 1mg
  • MA 200/1.5Y.NS Moisture Analyzer
    Max: 200g | d: 0,1mg
  • Compliance with 21 CFR part 11 (584.40 GBP)
  • Additional fee for Radwag Extended SI (RES level: 0,1,2) module activation (418.80 GBP)
  • Calibration new precision balances / industrial scales 0kg-60kg (189.60 GBP)
  • Pcs.

    Add to basket
    We are near you.
    We have been manufacturing weighing devices since 1984.

    RADWAG UK Limited
    16 Macon Court, Herald Drive, Crewe, CW1 6EA
    +44 01625 569569


    Delivery only on the territory of United Kingdom
    Optional paid installation and training service available - please contact us for details and pricing.
    Code:
    WL-304-0043
    CE
    CE
    2 years warranty
    2 years warranty

    Datasheet

    Maximum capacity [Max]
    50 g
    Readability [d]
    1 mg
    Tare range
    -50 g
    Adjustment
    external
    Leveling system
    semi-automatic – LevelSENSING
    Display
    10″ graphic colour touchscreen
    Weighing chamber
    automatic
    Protection class
    IP 43
    Delivery components
    Moisture Analyzer, disposable pan x 10, cross-shaped holder, drying pan handle, weighing pan shield, fabric dust cover, power cord
    Display mode
    %D, %M, %R, g
    Communication interface
    2×USB-A, USB-C, RS 232 (COM3), HDMI, Ethernet, Wi-Fi®, Hotspot
    Power supply
    100 – 120 V AC 50/60 Hz or 200-240 V AC 50/60 Hz
    Power consumption
    6 W
    Operating temperature
    +10 – +40 °C
    Maximum sample weight
    50 g
    Heating module
    IR emitter
    Heating module power
    450 W
    Moisture content repeatability
    ±0,05 % (sample ~ 2 g), ±0,01 % (sample ~ 10 g)
    Moisture content readability
    0,001 %
    Drying temperature range
    Max 160 °C
    Mass indication control
    External mass standard 50g, 100g, 200g, depending on the type of moisture analyzer (additional equipment)
    Drying temperature control
    One- or two-point, depending on the drying profile (additional equipment: GT105k-12/Z Control Thermometer)
    Moisture content indication control
    Certified MV Test Set (additional equipment)
    Drying temperature adjustment
    Three-point adjustment at defined temperatures, programmable values (additional equipment: GT105k-12/Z Control Thermometer)
    Drying method
    4 drying profiles (standard, fast, step, mild)
    Finish mode
    4 modes (automatic, manual, time-defined, user-defined)
    Maximum sample height
    20 mm
    Weighing pan dimensions
    ø90, h= 8 mm
    Packaging dimensions W x D x H
    540×610×470 mm
    Net weight
    5,3 kg
    Gross weight
    13 kg
    Additional functions
    sample traceability

    Overview


    MA 50.5Y Moisture Analyzer

     

    Do you need a moisture analyzer that not only determines moisture or dry matter content with impressive accuracy, but also performs a digital audit, responds to your voice commands and sends you light messages?

     

    If so, learn about the functionality of the ELLIPSIS series of moisture analyzers:

     

     

     

     

     

    It Dries in 4 Ways

    The MA 5Y Moisture Analyzer offers you four drying profiles: standard, mild, fast and step.
     
    • The standard profile is the usually used drying profile. It allows the most accurate determination of moisture content.
       
    • The mild profile is designed for drying substances that are sensitive to the rapid heat released by the filaments, which heat full power at the initial stage of the process. An example of such a substance is leather
       
    • You will use the fast profile for samples with significant moisture content – of the order of even several tens of percent.  
       
    • The step profile will be ideal for drying substances that contain more than 15 percent moisture. Within this profile you can define any three temperatures. You will choose the values of temperature and duration of heating experimentally.

     

     

     

     

     

    What Drying Temperature Should I Set?

    The drying temperature determines the duration of the process. Its value depends on the type of material. If the temperature is too low, it will under-dry the sample and, consequently, unnecessarily increase the measurement time, while if it is too high, it will cause a burning effect on the material. The drying temperature for the traditional (oven) method is determined by industry or factory standards. If there are no standards, you will select the temperature experimentally.
     
    The maximum drying temperature on the MA 5Y Moisture Analyzer is 160°C.

     

     

     

     

    How Do I Finish the Drying? In 4 Ways

    The MA 5Y Moisture Analyzer has four finish modes: automatic, manual, time-defined and user-defined.
    • Automatic mode
      You define the moisture loss limit value in 60 seconds. When the moisture analyzer reaches the set criterion, the measurement will automatically end.  
       
    • Manual mode
      You terminate the measurement yourself by pressing the STOP button.
       
    • Time-defined mode
      The measurement ends after the set time has elapsed regardless of the result. You have up to 99 hours 59 minutes to manage!
       
    • User-defined mode
      When you remove all volatiles from a sample, by definition, its mass should be constant. During drying, the dry mass of the sample is defined by a condition: if the mass of the sample is stable within 1 mg during the observation time (Auto 1-5), we consider that all volatile components have been removed from its structure.

     

     

     

     

     

    Digital Weighing Auditor

    Ensures that your moisture analyzer is ready for use. What is more, it informs about the need for a moisture analyzer inspection or a periodic audit of the moisture analyzer’s accuracy and sensitivity.

     

     

     

    10-Inch Display

    Bigger screen – 10’’ – means more information in one place. On the new ELLIPSIS terminal, you will see up to 3 screens, widgets, graphs, statuses, notes, handy measurement history.

     

     

     

     

     

    Uncompromising User Verification

    For the first time ever, the moisture analyzer will verify the user using a password, RFID card, fingerprint reader, facial recognition mechanism, or any combination of these.

     

     

     

     

     

    Ambient Light – an Innovative Way for the Moisture Analyzer to Communicate With the User

    The light colour will tell you about status, process results, procedures, or alerts.

     

     

     

     

     

    Hotspot

    Use the moisture analyzer on any device connected to it. This can be a smartphone, tablet, or computer.

     

     

     

    RFID

    ELLIPSIS works with RFID tags* that can be used to tag your products or formulation ingredients in the database. They can also be used to identify a user.

     

     

     

     

     

    Live Note – Note Down Your Conclusions

    It’s the first time you can add a text memo to a measurement series or procedure report. These can be, for example, assumptions, reminders or questions that can be returned to at any time.

     

     

     

     

     

    Handy Library

    This is where a series of recent measurements or a report ready for digital signature, according to 21 CFR Part 11, can be found.

     

     

     

     

    Widgets

    Prepared to display what you need most at any given moment.

     

     

     

     

    21 CFR Part 11

    Do you work in the pharmaceutical industry? Do you need a digital signature?
     
    We are offering the moisture analyzer which as a standalone fully meets the requirements of 21 CFR Part 11/ EU GMP Annex 11.
    • Full data security
    • Electronic signatures
    • Validations
    • Electronic records
    • Audit trails

     

     

     

    Modbus TCP/IP

    is one of the most popular communication protocols in industrial automation. Implementing it in the 5Y balances and CY10 scales makes the devices of this series friendly to automation integrators. 
     
    A balance or scale in a Modbus system is a slave device and performs tasks specified by the master device which is most often a PLC. Protocol Modbus TCP/IP in the 5Y balances and C10 scales enables, among other things, the realization of mass readout, zeroing and tare functions of the device, as well as tare settings and MIN and MAX thresholds.


     

     

     

     

    * RFID ISO/IEC 14443 Type A, 13.56 MHz