Datasheet
E1 Calibration Range
5 – 10 kg
E2 Calibration Range
1 – 10 kg
F1 Calibration Range
500 g – 10 kg
F2 Calibration Range
100 g – 10 kg
M1 Calibration Range
100 g – 10 kg
M2 Calibration Range
100 g – 10 kg
Maximum capacity [Max]
10,2 kg
Standard repeatability [Max]
0,4 mg
Standard repeatability [5% Max]
0,35 mg
Permissible repeatability
0,7 mg
Eccentricity (tested load)
2d / 1 mm
Electric compensation range
-100 g – +200 g
Adjustment
internal (automatic)
Display
10″ graphic colour touchscreen
Supplementary weights
internal (automatic)
Supplementary weights internal
half automatic
Supplementary weights external
300 g, 200 g
Communication interface
2×USB-A, USB-C, RS 232 (COM3), HDMI, Ethernet, Wi-Fi®, Hotspot
Operating temperature
+15 – +30 °C
Operating temperature change rate
±0,5 °C / 12 h (±0,3 °C / 1 h)
Relative humidity
40% – 60%
Relative humidity change rate
±5% / 4 h
Weighing pan dimensions
ø190 (ø300) mm
Weighing device dimensions
455×300×380 mm
Controling device dimensions
249×170×72 mm
Packaging dimensions W x D x H
1160×650×700 mm
Chamber packaging size
960×920×735 mm